Milk Fat Analyzer LAMA-B10

Milk Fat Analyzer LAMA-B10


Milk Fat Analyzer LAMA-B10 based on Soxhlet extraction principle, determines fat and SNF (Solids Non Fat) content in milk by (gravimetric) weight method. Integrated with processes such as soaking, extraction, leaching, heating, condensation and solvent recovery by means of alkali solvent and oil chromatographic fat separation. Batch handling sample volume and automated fat extraction method ensures safe and effective sample handling. This fully automated milk fat analyzer delivers fast and safe determination of fat constituents in milk and milk powder.





Test rangemilk fat ≥ 0.3 %
Sample sizeSolid: 1 to 10 g, liquid: 3 to 10 ml
Extraction time120 min
PrecisionFat content ≤ 5%, ≤ 0.1 g / 100 g, Fat content of 5 to 15%, ≤ 0.2 g / 100 g, Fat content ≥ 15%, ≤ 0.3 g / 100 g
Reproducibility≤ 0.4%
Capacity per batch6 samples
Solvent volume70 ml
Solvent recovery≥ 80 %
Temperature rangeRoom temperature to 200°C
Working temperature requirementCondensate temperature >4 °C, <26 °C
Temperature regulator accuracy± 0.5 °C
Control mode7 inch ultra-fine color touch screen LCD
Pumping bottle70 ml glass bottle
Extraction cupAluminium cup
Temperature protection systemIntelligent insulation control, with alarm
Heating methodAluminum alloy heating
Power supply220 V ± 10 %, 50 to 60 Hz
Power consumption500 W
Dimension (W × L × H)600 × 400 × 640 mm
Weight32 kg
  • Soxhlet extraction using alkali solvent like diethyl ether and petroleum ether
  • Fat cell breaking mechanism and oil chromatographic separation
  • Acid-base hydrolysis method to quantify weight of extract
  • Microcomputer control system with LCD touch screen display
  • Efficient solvent reflux recovery system by distillation
  • Equipped with six extraction positions
  • Unique air insulation for temperature heating protection and constant temperature
  • Integral aluminum alloy heating gives quick, uniform, stable and safe temperature rise
  • Core technology
  • Condenser tube integrated with oil chromatographic separation tube
  • Liposuction extraction bottle of glass for sample storage and sample stratification
  • Aluminium cup as receiving cartridge for recovery of fats
Used in testing milk and milk powder on different parameters for quality and testing, research purpose across milk packing and production testing plants, food and confectionary factories.