Download Catalog for Table top spectrophotometer LTS-A13
www.labtron.website | info@labtron.website
Transmitted spectrophotometer LTS-A13 with table top design is used to measure the transmittance, absorbency, concentration as well as reflectance of the material. Equipped with SCS system for best measurement repeatability & accurate measurement of surface color of material. The double grating & double optical path splitting system & ETC (every test calibration technology) produces 0.01% resolution, color balance intensity across visible spectrum & highly accurate measurements.
| Illumination system | d/0 (Diffused lighting, 0° observation viewing), SCS optical engine (light splitting and integration system), ETC (every time calibration technology), SCI (specular reflection included)/ SCE (specular reflection excluded) simultaneous measurement |
| Size of integrated sphere | 40 mm, alvan diffused reflection surface coating |
| Illumination light source | CLED |
| Sensor | Dual light path sensor array |
| Light source | A, C, D50, D65, D75, F1, F2, F3, F4, F5, F6, F7, F8, F9, F10, F11, F12, DLF, TL183, TL184, NBF, U30, CWF |
| Wavelength range | 400-700 nm |
| Wavelength interval | 10 nm |
| Half spectral width | 5 nm |
| Reflectivity range & resolution | 0 to 200%, 0.01% |
| Observation Angle | 2°/10° |
| Measurement time interval | 0.5 s |
| Measurement time | 2 s |
| Measuring aperture | 25.5*10 mm (cuvette support) |
| Color space | CIE-L*a*b, L*C*h, L*u*v, XYZ, Yxy, Transmittance, Hunterlab, Munsell Mi, CMYK |
| Color difference formula | ΔE*ab,ΔE*CH,ΔE*uv,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*94,ΔE*00,ΔEab(Hunter),555 color classification |
| Other colorimetric indices | WI (whiteness) (ASTM E313-10,ASTM E313-73,CIE/ISO, AATCC, Hunter, Taube Berger,Ganz, Stensby) YI (yellowness) (ASTM D1925,ASTM E313-00,ASTM E313-73), Tint(ASTM E313,CIE,Ganz) Metamerism index Milm, staining fastness, color fastness, APHA, Astm color, ISO illuminance, saybolt |
| Repeatability | Transmittance: standard deviation within 0.08% color values:ΔE*ab<=0.015 (After calibration, standard deviation of 30 measurements on test white board, 5 second intervals),Maximum:0.03 |
| Work temperature range | 0-45°C, relative humidity 80% or below (at 35°C), no condensation |
| Storage temperature range | 25℃ to 55℃ relative humidity 80% or below(at 35°C ),no condensation |
| Data being displayed | Sample's color values,color difference values/graph,pass/fail results,color error tendency,color value simulation, color difference values, display measurement area,history data color simulation,manual input standard sample, generate measurement report |
| Interface | USB |
| Data storage | Mass storage (on PC) |
| Light source longevity | 5 years, 1.5 million tests |
| Dimension (L*W*H) | 475 × 340 × 150 mm |
| Weight | 7 kg |