Overview
ICP Spectrometer LICP-B10 is an instrument which measures elements in ppb levels with least interference and reduces the time taken for analysis by detecting 5 to 6 elements in 1 min. Excellent analytical precision, high optical resolution, fully automated, wide wavelength range with 2400 line grating, powerful analysis software, high performance solid-state RF system, high accuracy etc. are some of the characteristic features. The equipment consist of a Scott-type spray chamber, quartz torch, spectrophotometer with Czerny-Turner monochromator, detector with photomultiplier tube and data processing system which allows multi-element measurement along with high resolution. It finds applications in environmental, food safety, metallurgy, pharmaceutical etc. fields.
Specifications :
| Monochromator Specifications |
| Optical type |
Czerny turner |
| Resolution |
≤ 0.015nm (3600 line grating) ≤ 0.030nm (2400 line grating) |
| Focal length |
1000 mm |
| Grating specifications |
huge holographic grating with 3600 L/mil or 2400 L/mil and 80 mm × 100 mm of ruling area |
| Wavelength range |
195 to 500 nm for 3600 line grating 195 to 800 nm for 2400 line grating |
|
| Solid State Power Specifications |
| Frequency |
27.12 MHz Frequency stability: < 0.05% |
| Spray chamber |
Scott double pass spray chamber |
| Output power |
800 W to 1600 W, adjustable with power efficiency more than 65% |
| Output power stability |
≤ 0.05% |
| Induction coil |
25 mm × 3 ID (ID-internal diameter), equipped with three concentric quartz torch tubes 35 mm ED (ED-external diameter) |
|
| Technical Specifications |
| Suitable sample content range |
Liquid sample: 0.01 ppm to several thousand ppm Solid or power sample: 0.001% to 70% |
| Repeatability |
short-term stability- RSD ≤ 1.5% Long-term stability: RSD ≤ 2% |
| Test speed |
5 to 8 elements / min |
| Limits of detection (LOD, µg/L) for most elements |
1ppb to 10ppb |
|
| Control Circuit |
| Photomultiplier tube specification |
R212 / R928 |
| Negative high voltage |
-50 to -1000 V |
| Circuit measuring range |
(10-12 to 10 -4) A |
| Signal acquisition |
VF conversion |
|
| Working Environment |
| Temperature for storage and transportation |
15°C to 25°C |
| Humidity for storage and transportation |
≤ 70% |
| Power Stability |
220 ± 10V; 50 to 60 V |
| Humidity |
≤ 70% |
| Temperature |
15°C to 30°C |
|
| Other Details |
| Packaging Dimension and Gross Weight (4 case) |
1st case: 188 × 111 × 105 / 348 kg 2nd case: 60 × 60 × 55 / 25 kg 3rd case: 59 × 67 × 97 / 80 kg 4th case: 38.5 × 38.5 × 63 / 47 kg |
|
Features :
- - Highly advance optical design
- - Holographic grating for minimum scattering of light
- - Correction of spectral interference to improvise detection
- - System controlled torch movement for optimized position
- - System controlled mass flow controller for carrier gas, auxiliary gas and plasma gas
- - 27.12 MHz RF generator for efficient ionization
- - Auto-coupling system
- - Peristaltic pump for stable flow rate and drainage of sample
- - Argon gas cylinder last up to 8 hours
- - Auto-sampler can be provided
- - Software Details
- - Allows selection of multiple elements and parameters for analysis
- - Faster Calibration function
- - Correction of background interference
- - Report generation with single button press
- - Many export options
- - Real-time graphical display of instrument status
- - Plasma ignition with one button press
- - Large database to select spectral lines according to requirement