Overview
Analytical HPLC-A7835 system is equipped with a flow rate range of 0.0001 to 10 ml per minute. Integrated with a single and dual wavelength UV-Vis detector with a range of 190 to 800 nm that analyzes sample’s absorption at different wavelengths and a high pressure pump providing continuous consistent flow. Built in column heater provides reproducible retention time.
Specifications :
| Pump Specifications |
| Model No |
HPP-B1 |
| Flow rate range |
0.0001 to 10 ml / min |
| Maximum pressure |
42 MPa , 0 to 5 ml / min |
| Column size |
4.6 × 250 mm |
| Sample size |
5 μm |
| Pressure pulsation |
≤ 1.5 % at 10 MPa |
| Flow rate accuracy |
± 0.2 % at 1 ml / min |
| Flow rate repeatability % |
≤ 0.15 |
| Gradient accuracy % |
± 2 |
| Gradient repeatability % |
≤ 0.2 |
| Dimension |
450 × 260 × 190 mm |
| Power supply |
90 V to 240 V |
| Weight |
11.2 kg |
|
| Detector Specifications |
| Model No |
HPD-B2 |
| Wavelength range |
190 to 800 nm |
| Wavelength accuracy |
± 0.75 nm |
| Wavelength repeatability |
0.2 nm |
| Band width |
8 nm |
| Optical path |
10 nm |
| Deuterium lamp |
190 to 400 nm |
| Tungsten lamp |
400 to 800 nm |
| Baseline noise ( Static ) Au |
≤ ± 0.5 × 10 -5 |
| Baseline drift ( Static ) Au / h |
≤ 1.5 x 10 -4 |
| Minimum detection limit |
≤ 4 x 10 -8 |
| Power |
100 W |
| Power supply |
90 V / 50 to 60 Hz |
| Dimension |
460 × 260 × 190 mm |
| Weight |
9.2 kg |
|
Features :
- - Compression compensation ensuring gradient repeatability
- - Low pressure gradient
- - Detector ( Deuterium or tungsten lamp )
- - Simultaneous working of deuterium and tungsten lamp
- - Automatic sample injector ( Autosampler )
- - External pump control by panel or computer software
- - Auto save last detected records by the detector
- - 10 point calibration pump for linear pressure control
- - Error alarm function for over pressure ,under pressure and leakage